Surface-distributed low-frequency asynchronous stimulation delays fatigue of stimulated muscles.

No Thumbnail Available
Identifiers
Publication date
2013-12
Advisors
Journal Title
Journal ISSN
Volume Title
Publisher
Citations
Google Scholar
Export
Research Projects
Organizational Units
Journal Issue
Abstract
One important reason why functional electrical stimulation (FES) has not gained widespread clinical use is the limitation imposed by rapid muscle fatigue due to non-physiological activation of the stimulated muscles. We aimed to show that asynchronous low-pulse-rate (LPR) electrical stimulation applied by multipad surface electrodes greatly postpones the occurrence of muscle fatigue compared with conventional stimulation (high pulse rate, HPR). We compared the produced force vs. time of the forearm muscles responsible for finger flexion in 2 stimulation protocols, LPR (fL = 10 Hz) and HPR (fH = 40 Hz). Results: Surface-distributed low-frequency asynchronous stimulation (sDLFAS) doubles the time interval before the onset of fatigue (104 ± 80%) compared with conventional synchronous stimulation. Combining the performance of multipad electrodes (increased selectivity and facilitated positioning) with sDLFAS (decreased fatigue) can improve many FES applications in both the lower and upper extremities.
Description
Citation
Maneski , L Z P , Malešević , N M , Savić , A M , Keller , T & Popović , D B 2013 , ' Surface-distributed low-frequency asynchronous stimulation delays fatigue of stimulated muscles. ' , Muscle and Nerve , vol. 48 , no. 6 , pp. 930-937 . https://doi.org/10.1002/mus.23840